Difference between revisions of "October 07 2009 Meeting: VT-NAMIC Data review summary"

From NAMIC Wiki
Jump to: navigation, search
(Created page with 'We made the following observations about the data at the first time point: * In some sobjects the FOV is axis-aligned, while in others it is tilted. These directions were not pr…')
 
 
(8 intermediate revisions by one other user not shown)
Line 1: Line 1:
 +
== General comments ==
 +
 
We made the following observations about the data at the first time point:
 
We made the following observations about the data at the first time point:
  
Line 6: Line 8:
  
 
Other observations with illustrations:
 
Other observations with illustrations:
 +
 +
== Acquisition-related issues ==
 +
 +
 +
{|
 +
| '''Angio sequence (right name?) has apparently been used for some (7 out of 10) subjects''' CLW: This is normal T1w blood flow artifact, but it is pronounced, perhaps because of the small brain size.
 +
{|
 +
|
 +
|<gallery perrow="3" heights="300px" widths="800px">
 +
Image:calvin-angio.jpg|'''Calvin axial showing vessel and volume rendering of vasculature'''
 +
</gallery>
 +
|}
 +
|}
  
 
{|
 
{|
| '''The issue with noise in Tommy and Louis resolved: we do not know how the NIFTI images we used before for Tommy and Louis were obtained -- the DICOM data has much better noise characteristics'''
+
| '''Fat shift artifact significantly obstructs the cortex in the occipital lobe for all subjects'''
 
|
 
|
 
{|
 
{|
|<gallery perrow="2" heights="200px" widths="300px">
+
|<gallery perrow="3" heights="200px" widths="300px">
Image:tommy-dicom.jpg|'''Tommy, DICOM'''
+
Image:calvin-artefact1.jpg|'''Calvin occipital lobe'''
Image:louis-dicom.jpg|'''Louis, DICOM'''
+
Image:calvin-artefact2.jpg|'''Calvin'''
 +
Image:calvin-artefact3.jpg|'''Calvin'''
 
</gallery>
 
</gallery>
 +
|}
 
|}
 
|}
  
 +
{|
 +
| '''Ringing artifacts for some (few) subjects'''
 +
|
 +
{|
 +
|<gallery perrow="3" heights="200px" widths="300px">
 +
Image:calvin-ringing.jpg|'''Calvin ringing artifact'''
 +
</gallery>
 +
|}
 
|}
 
|}
  
 
{|
 
{|
| '''Angio sequence (right name?) has apparently been used for some (7 out of 10) subjects'''
+
| '''Zero filling artifact'''
 
|
 
|
 
{|
 
{|
|<gallery perrow="3" heights="300px" widths="800px">
+
|<gallery perrow="3" heights="200px" widths="400px">
Image:calvin-angio.jpg|'''Calvin axial showing vessel and volume rendering of vasculature'''
+
Image:calvin-zero_filling.jpg|'''Calvin time point 2'''
 +
</gallery>
 +
|}
 +
|}
 +
 
 +
{|
 +
| '''Susceptibility artifacts'''
 +
|
 +
{|
 +
|<gallery perrow="3" heights="200px" widths="400px">
 +
Image:calvin-susceptibility.jpg|'''Calvin time point 1'''
 +
Image:gucci-susceptibility.jpg|'''Gucci time point 1'''
 
</gallery>
 
</gallery>
 
|}
 
|}
 
|}
 
|}
 +
 +
 +
== Processing-related issues ==
  
 
{|
 
{|
| '''Strange artifact is observed in the occipital lobe for all subjects. Ron, Sandy, Sota Oguro (SPL radiologist) cannot explain the origin. See Calvin example (note the interpolation is turned off)'''
+
| '''The issue with noise in Tommy and Louis resolved: we do not know how the NIFTI images we used before for Tommy and Louis were obtained -- the DICOM data has much better noise characteristics'''  
 
|
 
|
 
{|
 
{|
|<gallery perrow="3" heights="200px" widths="300px">
+
|<gallery perrow="2" heights="200px" widths="300px">
Image:calvin-artefact1.jpg|'''Calvin occipital lobe'''
+
Image:tommy-dicom.jpg|'''Tommy, DICOM'''
Image:calvin-artefact2.jpg|'''Calvin'''
+
Image:louis-dicom.jpg|'''Louis, DICOM'''
Image:calvin-artefact3.jpg|'''Calvin'''
 
 
</gallery>
 
</gallery>
 
|}
 
|}
 +
 
|}
 
|}
  
 
{|  
 
{|  
| '''Significant mis-alignment of some subjects with the template subject after FLIRT registration is confirmed'''
+
| '''Significant mis-alignment of some subjects with the template subject after FLIRT registration for some subjects is confirmed''' Should be fixed in the new analysis.
 +
|
 
{|
 
{|
 
|<gallery perrow="3" heights="200px" widths="300px">
 
|<gallery perrow="3" heights="200px" widths="300px">
Image:ralph-mis-reg.jpg|'''Calvin occipital lobe'''
+
Image:ralph-flirt.jpg|'''Ralph/atlas post-FLIRT checkerboard'''
 +
Image:mark-flirt.jpg|'''Mark/atlas post-FLIRT overlay'''
 
</gallery>
 
</gallery>
 
|}
 
|}
 
|}
 
|}
 +
 +
 +
== Biology-related issues ==
  
 
{|
 
{|
| '''Ringing artifacts for some (few) subjects'''
+
| '''Valentino has structure not consistent with the normal anatomy''' ''Ron: this is likely a cyst.''
 
|
 
|
 +
{|
 +
|<gallery perrow="2" heights="200px" widths="300px">
 +
Image:valentino-issue1.jpg|'''Valentino unknown structure, Fast marching segmentation'''
 +
Image:valentino-issue2.jpg|''''''
 +
Image:valentino-issue3.jpg|''''''
 +
Image:valentino-issue4.jpg|'''Valentino unknown structure, surface model'''
 +
</gallery>
 +
|}
 
|}
 
|}

Latest revision as of 13:43, 23 December 2009

Home < October 07 2009 Meeting: VT-NAMIC Data review summary

General comments

We made the following observations about the data at the first time point:

  • In some sobjects the FOV is axis-aligned, while in others it is tilted. These directions were not preserved in the NIFTI data we were previously using.
  • We have two SPGR scans of for each subject, but the scans are not perfectly aligned. Ron: possibly due to coil heating. We cannot right away use these images to average and get better SNR, need to check with imaging people. Not sure if discrepancy is the rigid one.
  • Noticeable bias field in the atlas is observed by Ron.

Other observations with illustrations:

Acquisition-related issues

Angio sequence (right name?) has apparently been used for some (7 out of 10) subjects CLW: This is normal T1w blood flow artifact, but it is pronounced, perhaps because of the small brain size.
Fat shift artifact significantly obstructs the cortex in the occipital lobe for all subjects
Ringing artifacts for some (few) subjects
Zero filling artifact
Susceptibility artifacts


Processing-related issues

The issue with noise in Tommy and Louis resolved: we do not know how the NIFTI images we used before for Tommy and Louis were obtained -- the DICOM data has much better noise characteristics
Significant mis-alignment of some subjects with the template subject after FLIRT registration for some subjects is confirmed Should be fixed in the new analysis.


Biology-related issues

Valentino has structure not consistent with the normal anatomy Ron: this is likely a cyst.