Difference between revisions of "NANOMA:Artifacts:Experiments"

From NAMIC Wiki
Jump to: navigation, search
 
Line 35: Line 35:
 
**Frequency : 256
 
**Frequency : 256
 
**Phase : 160
 
**Phase : 160
 
  
 
==DICOM Images==
 
==DICOM Images==

Latest revision as of 15:55, 8 July 2009

Home < NANOMA:Artifacts:Experiments

Image artifact experiments for magnetic particles conducted at BWH on June 20, 2009

  • Objective: To characterize the image artifacts created by magnetic particles. We have imaged the magnetic particles for different pulse sequences with variations in echo type, echo time, voxel size and flip angle. We have also used particles of varying diameter and material.
  • Where : Brigham and Women's Hospital
  • Scanner : GE Signa 3.0T
  • When : June 20, 2009
  • Participants : Jayender Jagadeesan (BWH), Tim Wortmann (University of Oldenburg)

Sequences used

  • Gradient Echo (GRE)
    • TE : Minimum Full
    • TR : 22
    • Flip Angle : 30
    • Bandwidth : 15.63
    • FOV : 16
    • Slice thickness : 3
    • Frequency : 256
    • Phase : 160
  • Spin Echo (SE)
    • TE : Min Full
    • TR : 500
    • Bandwidth : 31.25
    • FOV : 16
    • Slice thickness : 3
    • Frequency : 256
    • Phase : 160
  • Single Shot Fast Spin Echo (SSFSE)
    • TE : 80
    • TR : Min
    • Bandwidth : 83.33
    • FOV : 16
    • Slice thickness : 3
    • Frequency : 256
    • Phase : 160

DICOM Images

_

_

_

_

_

_

_

_

_

_

_

_

Photographs

  • Slice view of agar