CTSC:MGHresources:Integrated MR-PET imaging system

From NAMIC Wiki
Jump to: navigation, search
Home < CTSC:MGHresources:Integrated MR-PET imaging system

Back to CTSC:MGHresources

The bay 8 laboratory contains the first human MR-PET system installed in the US. It is an insert PET camera (Siemens Healthcare) employing APD photodiode detectors and is capable of simultaneous recording of the MR and PET image in the brain. The BrainPET is a 3D dedicated-brain scanner designed to operate inside the bore of the Siemens Medical MAGNETOM Trio, a TIM system 3T MR scanner. The combined MR-PET scanner is the first system capable of simultaneously imaging the human brain in one bed position.

The PET system consists of the following major components:

  • PET gantry with 32 detector cassettes

The gantry physical ID and OD are 36 cm and 60 cm, respectively. The axial FOV is 19.25 cm and the transaxial FOV is ~30 cm. Each of the 32 PET detector modules consists of six 12×12 LSO crystal arrays read out by a 3×3 array of APDs (Hamamatsu 8664-55, Japan). The individual crystal size is 2.5×2.5×20 mm3. The APD signals are processed and shaped by an on-board charge-sensitive preamplifiers and pole-zero circuit. The signals from the PET modules are taken from the gantry via 10-meter long cables to the acquisition electronics module.

  • gantry handling device with cable handling function

The gantry handling device performs multiple functions: supports the PET detector gantry, supports signal cables and cooling tubes connected to the gantry, incorporates a sliding rail platform to allow transaxial positioning of the gantry, incorporates an articulated cable race that allows axial positioning of the gantry. An internal control system regulates the gantry internal temperature.

  • cooling system, acquisition electronics module

The air/water based cooling system includes a compressor and a chiller. The acquisition electronics module includes an embedded command and control computer to control the various hardware components such as the detector array, data collection circuits, etc. The embedded computer also acts as a server on a dedicated Ethernet network.

  • processing reconstruction station and power supply cabinet.

Also connected to this network is the processing and reconstruction station (PC, Intel Xeon X5355 @2.66 GHz quad processor; OS: Windows Server 2003; Memory: 16 GB) that serves as the primary user interface to the BrainPET subsystem. Running on this workstation is the acquisition workplace software application that supports acquisition, reconstruction, data management, and workflow definition.